An Agilent 1200 HPLC system (Agilent, USA) with electrospray ionization coupled to an Agilent 6460 triple quadruple mass spectrometer (LC-ESI-MS/MS) was used to confirm the structures of the GSLs. The HPLC conditions were the same as those mentioned in the previous section, and the mass conditions are listed in Table 6. Compounds with m/z = 75 and featuring [M-G-H] molecular ion peaks were selected as candidate compounds [90 (link), 91 ]. The positive ion peak, such as [M-G + H]+, was used for identification [51 (link), 92 (link)]. For MS/MS conditions, the fragmentor voltage was optimized by approximately 1/3 of the molecular weight, and the collision energy number was set to approximately 1/15 of the given molecular weight.

LC-MS/MS conditions

ItemsParameters
Gas Temp300 °C
Gas Flow10 L/min
Nebulizer45 psi
Sheath Gas Temp350 °C
Sheath Gas Flow11 L/min
Capillary3000 V(+)/3000 V(−)
Nozzle Voltage500 V(+)/500 V(−)