Ultra-high-performance liquid chromatography coupled with Q-Orbitrap (UPLC-Q-Orbitrap HRMS, Thermo Fisher Scientific, Waltham, MA, USA) was previously applied to speculate the components of SCE [7 (link)]. Standard identification of abietic acid (HPLC ≥ 98%, Solarbio, Beijing, China) is essential for the rigor of this present study. The sample (200 mg) was dissolved in methanol solution (methanol: water, 8:2, V/V; 1 mL) before centrifugation at 20,000× g for 10 min. Subsequently, impurities were removed using a 0.22-µm PES membrane. Abietic acid was diluted with methanol and sonicated for 5 min to obtain a standard stock solution with a concentration of 20 µg/mL. The mass spectrometry was performed via ESI with positive and negative ion switching scanning. For chromatography, it was accomplished according to gradients of aqueous and organic phases. More details can be obtained in our previous study [8 (link)]. The quantification of AA was determined using peak area calculation.