The dissolution experiment of neat CorA and CorA-ASD formulations were performed using the biphasic dissolution apparatus BiPHa+ [31 (link),32 (link)]. The experimental setup for the aqueous phase was identical to the monophasic setup (Section 2.4.1). Additionally, 50 mL of 1-decanol were added simultaneously to the pH shift after 20 min from pH 3.0 to pH 4.8 covering the aqueous phase [23 (link)]. The concentration profiles of both phases were measured online continuously, with an Agilent 8454 UV-Vis spectrophotometer, similar to Section 2.4.1. Figure 1 shows the experimental setup of the biphasic dissolution.