Single-particle tracking photoactivated localization microscopy of MCF and MDCK cells transfected with Cav1-mEos2 was performed on the Roper Scientific total internal reflection fluorescence (TIRF) microscope equipped with an iLas2 double-laser illuminator (Roper Technologies), a CFI Apo TIRF 100× (1.49 NA) objective (Nikon), and an Evolve512 delta EMCCD camera (Photometrics). Images were acquired using Metamorph software (version 7.78; Molecular Devices) at 50 Hz, and 16,000 frames were acquired per cell. A 405-nm laser was used to photoconvert mEos2, with simultaneous 561-nm exposure to excite the photoconverted mEos2. For stochastic photoconversion of mEos2 molecules, a low amount (3–5%) of 405-nm laser and 75–80% of 561-nm laser was used. Data analysis was performed as previously described (Bademosi et al., 2017 (link); Kasula et al., 2016 (link)) using PALM-Tracer, a plugin in Metamorph software (Molecular Devices).