Using a scalpel, a longitudinal piece of each organ was harvested and weighed. Samples weighed between 0.01 g and 0.26 g. After weighing, the samples were added to a centrifuge tube. 500 μl of Aqua Regia (1 : 3 nitric acid : hydrochloric acid) was added to begin digestion. A glass rod was used to break down the sample into small pieces. The samples were allowed to digest for 1–7 days. Samples were diluted to 5% Aqua Regia with purified DI water. Samples were mixed by vortexing and then filtered through a 1.0 μm Nylon syringe filter before testing. The samples were run on a Perkins Elmer Avio 200 using a 5-point standard curve created from Si, Ag, and Au standards (LabChem). The instrument was blanked using a 5% Aqua Regia and purified DI water solution. The concentrations of elements in each sample were related to the weight of digested samples.