The DIMS analysis for the evaluation of metabolite profile in tear strip samples was performed using an Ultra Performance Liquid Chromatography Tandem Quadrupole Mass Spectrometry (UPLC/MS/MS) system (Acquity UPLC I-Class coupled to a Xevo TQD, Waters Corp., Manchester, UK). Mass spectra were acquired in positive electrospray ionization, using multiple reaction monitoring (MRM) as acquisition mode. Data are processed by using MassLynx V4.1 and NeoLynx Software (Waters Corp.). 10 μL were injected into the ion source and the run time was 1.1 minutes, injection-to-injection. The specific mass spectrometer setting conditions are reported in our previous works [37 (link),38 (link)]. A detailed list of analyzed metabolites, their ionization source settings, and their abbreviations as used in the text are available in Supplementary Table S5. The mass spectrometer ionization source settings were optimized for maximum ion yields for each analyte.
Free full text: Click here