PET emission data were acquired on a Siemens ECAT HR+PET scanners as previously described [35 (link)]. Briefly, participants were positioned in the scanner approximately 35 min after [11C]PiB injection (15 mCi). A 10 min transmission scan was acquired using rotating 68Ge/68Ga rod sources to correct for photon attenuation, followed by a 20 min emission scan (4 × 5 min frames) beginning 50 min after [11C]PiB injection. PET emission data were reconstructed using filtered back-projection (Direct Inverse FourierTransform) into a 128 × 128 × 63 matrix with voxel sizes of 2.06 × 2.06 × 2.43 mm3. Images were filtered with a 3mm Hann window.