Unless contraindicated, children underwent magnetic resonance imaging (MRI) using a Siemens Prisma® (Siemens Medical Solutions, Malvern, PA), General Electric Discovery MR750 (GE Healthcare, Chicago, IL), or Philips Achieva (Koninklijke Philips, Amsterdam, Netherlands) scanner. Standard adult coils provided high resolution images harmonized using the following settings: T1-weighted: MP-RAGE with 256×256 matrix size, 176 slices, 1 mm resolution in all axes, TR=6.31 ms, TE=2500 ms and TI=1060 ms; flip angle=8 deg; 2-fold parallel imaging; multiband off.12 (link) Cortical thickness, area, and volume were calculated using a Freesurfer-based pipeline13 (link) using the Destrieux atlas,14 (link) which parcellated 74 cortical and subcortical brain regions/hemisphere. Cortical effect size maps were generated using Nilearn.15 (link)