The sequence was simulated by Bloch equations to calculate the transverse magnetization as a function of all the protocol and tissue parameters in order to construct the LUT corrections for both the AIF and myocardial imaging protocols. Input to the LUT was the normalized signal SR/PD, where PD used the FLASH protocol and SR used either a b-SSFP or FLASH protocol. LUT were validated by phantom measurement by comparing the estimates of [Gd] after LUT correction with the known [Gd] using least squares fitting. A set of gadolinium doped saline phantoms were constructed at concentrations up to 10 mmol/L using both Gadoterate meglumine (Dotarem, Guerbet LLC) and Gadobutrol (Gadavist, Bayer Healthcare). LUT estimates of [Gd] vs known [Gd] were calculated with and without T2* correction. The phantom T1 values were measured using an inversion recovery GRE sequence at multiple inversion times (TI) with TR = 10s such that the longitudinal magnetization was fully relaxed after each RF excitation, and T1 was estimated by 3-parameter fitting to the mono-exponential inversion recovery S = A-Bexp(-TI/T1). The phantom T2 values were measured using a spin echo sequence (TR = 10s) with varying echo times (TE) and using T2 estimates from a 2-parameter fit to the mono-exponential decay curve, S = Aexp(-TE/T2). The coefficients for relaxivity rates (r1 and r2) were calculated from the T1 and T2 measurements vs known [Gd] using linear fitting, i.e., R1 = R10 + r1[Gd] with R1 = 1/T1.