MRI was performed using a dedicated knee coil in a 1.5T system (Philips, Medical Systems, Best, The Netherlands). Our standardised scanning protocol consisted of1 (link) Coronal proton density (PD) turbo spin echo (TSE), repetition time (TR)/echo time (TE) 2335/35 ms; echo train length (ETL) 6,2 (link) coronal frequency selective fat-suppressed PD TSE (TR/TE 2334/35 ms; ETL 6, 3 mm slice thickness);3 (link) sagittal PD TSE (TR/TE 2338/35; ETL 6; 3.5 mm slice thickness);4 (link) sagittal frequency selective fat-suppressed T1-weighted three-dimensional gradient echo sequence (TR/TE 11/5.5; 25° flip angle; 150 mm field of view, 272×512 acquisition matrix, 2 mm slice thickness with a 1 mm overlap between images);5 (link) axial frequency selective fat-suppressed PD TSE (TR/TE 3225/15; ETL 6, 4 mm slice thickness). In all TSE sequences, we used a 150–160 mm field of view and a 304×512 acquisition matrix. Total acquisition time, including the initial survey sequence, was 30 min.