ASL was performed using pseudocontinuous labeling with a 3D stack-of-spirals fast spin echo readout; this reflects the GE product ASL sequence. PCASL-specific parameters included a labeling duration of 1,450 ms and post labeling delay of 2025 ms with 3D spiral readout parameters as follows: spiral interleaves = 8; points per spiral = 512; slices = 36; slice thickness 4.0–4.2 mm; FOV = 24–26 cm; in-plane resolution = 3.64–4.53 mm2; bandwidth = 62.5 kHz; TE = 9.5–10.5 ms; TR = 4,800–4,847 ms; NEX = 3; and scan time = 4 min 32 s–4 min 42 s.
Pseudocontinuous ASL protocol on 3T MRI
ASL was performed using pseudocontinuous labeling with a 3D stack-of-spirals fast spin echo readout; this reflects the GE product ASL sequence. PCASL-specific parameters included a labeling duration of 1,450 ms and post labeling delay of 2025 ms with 3D spiral readout parameters as follows: spiral interleaves = 8; points per spiral = 512; slices = 36; slice thickness 4.0–4.2 mm; FOV = 24–26 cm; in-plane resolution = 3.64–4.53 mm2; bandwidth = 62.5 kHz; TE = 9.5–10.5 ms; TR = 4,800–4,847 ms; NEX = 3; and scan time = 4 min 32 s–4 min 42 s.
Corresponding Organization :
Other organizations : University of California System
Variable analysis
- Contrast agent: gadobenate dimeglumine (Bracco Diagnostics) or gadobutrol (Bayer AG)
- Contrast agent dose: 0.1 mmol/kg
- ASL signal
- MRI scanner: 3T Discovery 750 (GE Healthcare, Milwaukee Wisconsin)
- Brain coil: 8-channel brain array coil
- FSPGR imaging parameters: TE/TR = 3.0/6.9 ms, FA = 9°, FOV = 25 cm, matrix = 256 × 256, slices = 180, slice thickness = 1 mm, interslice gap 1 mm
- PCASL parameters: labeling duration = 1,450 ms, post labeling delay = 2025 ms, spiral interleaves = 8, points per spiral = 512, slices = 36, slice thickness = 4.0–4.2 mm, FOV = 24–26 cm, in-plane resolution = 3.64–4.53 mm^2, bandwidth = 62.5 kHz, TE = 9.5–10.5 ms, TR = 4,800–4,847 ms, NEX = 3, scan time = 4 min 32 s–4 min 42 s
- Positive control: Not specified
- Negative control: Not specified
Annotations
Based on most similar protocols
As authors may omit details in methods from publication, our AI will look for missing critical information across the 5 most similar protocols.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!