The detector voltage was set to 1350 V, and the electron energy was −70 V.
Comprehensive GC-GC-TOFMS Analysis Protocol
The detector voltage was set to 1350 V, and the electron energy was −70 V.
Corresponding Organization : State Key Laboratory of Food Science and Technology
Variable analysis
- Column configuration for GC×GC analysis (two columns coupled in tandem)
- Measured outcomes from the LECO Pegasus® 4D GC×GC-TOFMS system
- Helium carrier gas flow rate (1.0 mL/min)
- Injector temperature (250 °C)
- GC oven temperature program (initial 45 °C for 3 min, then ramped at 4 °C/min to 150 °C, held for 2 min, reaching 200 °C at 6 °C/min and 230 °C at 10 °C/min for 20 min)
- Secondary oven temperature (5 °C offset relative to primary oven)
- Modulator temperature (20 °C offset relative to primary oven)
- Modulation period (4 s)
- TOFMS acquisition rate (100 spectra/s)
- TOFMS mass scan range (50 m/z to 350 m/z)
- Detector voltage (1350 V)
- Electron energy (-70 V)
Annotations
Based on most similar protocols
As authors may omit details in methods from publication, our AI will look for missing critical information across the 5 most similar protocols.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.
Ready to get started?
Sign up for free.
Registration takes 20 seconds.
Available from any computer
No download required
Revolutionizing how scientists
search and build protocols!