We acquired imaging data at the Athinoula A. Martinos Center for Biomedical Imaging at MIT with a Siemens Magnetom Trio 3-tesla scanner with a 32-channel phased-array head coil. T1-weighted structural images were collected using the magnetization-prepared rapid acquisition gradient echo (MPRAGE) sequence (TR = 2530 ms; TE = 1.64–7.22 ms; TI = 1400 ms; flip angle = 7°; 1 × 1 × 1-mm3 isotropic voxels; matrix size: 256 × 256; 172 slices). Whole-brain diffusion-weighted images were collected with a spin-echo echo-planar sequence (TR = 8420 ms; TE = 84 ms; 2 × 2 × 2 mm3 isotropic voxels; matrix size: 128 × 128; 67 slices). This included 60 gradient orientations at b = 700 s/mm2 and 10 no-diffusion images (b = 0). Sixty-two volumes of eyes-open resting state data were collected with a 6 s TR. As with the diffusion images, the resting state matrix size was 128 × 128 × 67 with 2 × 2 × 2 mm3 isotropic voxels. T1 and diffusion data from these subjects were previously published (Cai et al., 2014 (link)). Resting state data were collected from the same subjects in the same MRI sessions.