PANI based SC-ISEs has improved potentiometric characteristics and short-term stability, while underneath the ion selective membrane an aqueous layer had been formed affecting the long-term stability. The formation of these layer was investigated by Pretsch group35 (link) introducing potentiometric aqueous layer test. This test was done to detect the potential drift upon the changing happened from an ALF solution (1.0 × 10−3 mol L−1) to the highly concentration interfering ion solution, Antazoline HCl (1.0 × 10−2 mol L−1) and back then to ALF solution (1.0 × 10−3 mol L−1). If the aqueous layer accumulated underneath the ion sensing membrane, potential drift would result from changing the aqueous layer ionic composition.