Metabolite Analysis by HPLC
Corresponding Organization : Universidad de Valladolid
Other organizations : HES-SO Valais-Wallis
Variable analysis
- Centrifugation speed (14,000 rpm)
- Centrifugation duration (5 minutes)
- Centrifugation temperature (4°C)
- Filtration method (Nylon syringe filters of 0.22 µm)
- Mobile phase (H2SO4 25 mM)
- Column temperature (75°C)
- Eluent flow rate (0.7 mL min-1)
- Concentration range (0.8–90 mM)
- Compound concentrations measured by HPLC
- Culture sample volume (2 mL)
- HPLC instrument (Alliance Waters HPLC)
- HPLC column (Aminex HPX-87H column, 7.8 mm × 300 mm)
- Positive control: Compounds obtained from Sigma-Aldrich (ReagentPlus, 99%)
Annotations
Based on most similar protocols
As authors may omit details in methods from publication, our AI will look for missing critical information across the 5 most similar protocols.
About PubCompare
Our mission is to provide scientists with the largest repository of trustworthy protocols and intelligent analytical tools, thereby offering them extensive information to design robust protocols aimed at minimizing the risk of failures.
We believe that the most crucial aspect is to grant scientists access to a wide range of reliable sources and new useful tools that surpass human capabilities.
However, we trust in allowing scientists to determine how to construct their own protocols based on this information, as they are the experts in their field.