Native MS was performed using a Q-Exactive HF quadrupole-Orbitrap mass spectrometer with the Ultra-High Mass Range (UHMR) research modifications (Thermo Fisher Scientific, Bremen). Instrumental parameters were used as previously described.12 (link),13 (link),15 (link),16 (link) Nano-electrospray ionization (ESI) was used to introduce the sample into the mass spectrometer with a capillary voltage of 1.1 kV and capillary temperature of 200 °C. Nanodiscs were analyzed from 2,000–25,000 m/z at a resolution setting of 15,000 and a trapping gas setting of 7. The source fragmentation was set to 50 V to aid in desolvation, and in-source trapping was set to 0 V. To analyze free peptides, the MS was operated from 500–5000 m/z with the resolution set to 120,000 and a trapping gas setting of 3. Source fragmentation was kept at 50 V to aid in desolvation.